Highlights | Specifications | To Order | Product Information | Home |
omegadyne.com |
THICK FILM | |||||||||||||||||
SERIES | PXM01/PXM02 Series | PXM103 Series | PXM41 Series | PXM78 Series | PXM80 Series | |||||||||||||
|
|
|
|
|
|
|||||||||||||
DESIGN FEATURES | High Accuracy Hermetically Sealed Transducer. FM/CSA approved. Intrinsically safe. On Amplified Outputs |
Flush diaphragm for slurries and viscous fluids. On Amplified Outputs |
FM/CSA intrinsically safe models available other dimensions, threads and wetted parts are available On Amplified Outputs |
Submersible level detectors with vented cable. Hermetically sealed in all stainless steel. On Amplified Outputs |
Differential pressure measurement ranging from ±100 m bar to ±350 bar with up to 414 bar line pressure. FM/CSA approved. Intrinsically safe. On Amplified Outputs |
|||||||||||||
PRESSURE CONNECTIONS Metric Adaptors Available |
M-MS33649-4 Female Metric and BSP threads available |
Flush Mount | Combination- 1/2 NPT External 1/4 NPT Internal |
Protective Nose cap included | 1/4 NPT 48 bar 1/8 NPT > 48 bar Metric Adaptors Available |
|||||||||||||
OUTPUTS AVAILABLE (FSO=Full Scale Output) |
3 mV/V ±.25% FSO, 0-5 Vdc ±10% adj 4-20 mA ±10% adj |
2 mV/V ±1% FSO, 0.5-5.5 Vdc 0.5-10.5 Vdc 4-20 mA |
3 mV/V 0.5-5.5 Vdc ±10% adj 4-20 mA ±10% adj |
3 mV/V 4-20 mA |
2 mV/V 0-5 Vdc 4-20 mA ±10% adj |
|||||||||||||
ACCURACY Linearity Hysteresis Repeatability |
0.05% FSO (PXM01)0.25% 0.15% FSO (PXM02) 0.05% FSO 0.05% FSO |
0.3 to 3.5 bar ±1% 4 to 350 bar ±0.25% >350 bar ±1% |
0.25% FSO Linearity, Hysteresis and Repeatability Combined |
0.25% FSO Linearity, Hysteresis and Repeatability Combined |
2 bar 0.5% FSO 2 bar 0.25% FSO Linearity, Hysteresis and Repeatability Combined |
|||||||||||||
ZERO BALANCE | ±1.0% FSO | ±1.5% FSO | +4 mA +10% -2% adj. | +4 mA | +4 mA +10% -2% adj. | |||||||||||||
EXCITATION Recommended Maximum |
10 V ac/dc 15 V ac/dc |
5 V ac/dc 10 V ac/dc |
10-40 Vdc 40 Vdc |
10-40 Vdc 40 Vdc |
12-40 Vdc 40 Vdc |
|||||||||||||
OPERATING TEMP Range
COMPENSATED TEMP |
-73 to +163°C -100 to +325°F
+17 to +71°C |
-54 to +93°C -65 to +200°F
0 to +55°C |
-45 to +121°C -50 to +250°F
+17 to +71°C |
-29 to +93°C -20 to +200°F
+17 to +71°C |
2 bar -34 to 93°C (-30 to 200°F) >2 bar -46 to 107°C (-50 to 225°F)
2 bar -1 to 54°C |
|||||||||||||
TEMP. EFFECTS Zero Span |
0.0018% FSO/°C (PXM01) 0.0045% FSO/°C (PXM02) 0.0027% FSO/°C (PXM01) 0.0054% FSO/°C (PXM02) |
0.018% FSO/°C 0.018% FSO/°C |
0.0081% FSO/°C 0.0054% FSO/°C |
0.0081% FSO/°C 0.0054% FSO/°C |
2 bar 0.022% FSO/°C >2 bar 0.018% FSO/°C 2 bar 0.022% FSO/°C >2 bar 0.009% FSO/°C |
|||||||||||||
WETTED PARTS | 17-4 PH Stainless Steel | Hastelloy C-276 or 17-4 PH Stainless Steel |
17-4 PH Stainless Steel with 316 Stainless Steel Diaphragm |
17-4 PH Stainless Steel | 2 bar 316 Stainless Steel >2 bar 17-4 PH Stainless Steel |
|||||||||||||
SAFE OVER-PRESSURE | 150% | 200% rated pressure | 150% rated pressure | 150% rated pressure | 138 bar or 150%, whichever is greater | |||||||||||||
BURST PRESSURE | 300% | 300% rated pressure | 300% rated pressure | 300% rated pressure | 414 bar minimum | |||||||||||||
ORDERING EXAMPLE PRICE |
PXM02M1-75V 368£ 3.5 bar sealed, mV/V output, (7/16)-20 Female thread PXM01 PXM02 |
PXM103UO-7AV 290£ 7 bar absolute, mV output transducer |
PXM41TO-70KGI 372£ 70 bar relative, 4-20 mA output pressure transmitter |
PXM78U7-2GI 561£ 2 bar relative, 4-20 mA output pressure transmitter |
PXM80B5-140mDI 852£ 140 mbar differential, 4-20 mA output transmitter |
|||||||||||||
For More Information Request Document # | #3605/#3606 (mV Output) | #3607 (mV Output) | #3608 (Current Output) | #3609 (Current Output) | #3610 (Current Output) |
Conversion Factors: |
1 dyne = 2.248x10-6 lb< | 1 lb = 44.48x104 dynes | 1 N=0.2248 lb | 1 lb = 4.448 N | 1 lb = 454 gm |
1 Bar = 14.504 psi | 1 psi = 6.894x10-2 Bar | 1 Kg/cm2 = 14.223 psi | 1 psi = 7.03x10-2 Kg/cm2 | 1 gm = 2.202x10-3 lb |
OMEGADYNE DELIVERS... | email: info@omegadyne.com |
Experienced Application Engineers When you need help in selecting the proper products for your application, call our qualified product specialist for assistance. Custom Design Services |
Repair and Calibration Services OMEGADYNE offers in-house mil-spec calibration and full repair services for all our products. Fast Delivery from Stock |
Highlights | Specification | To Order | Product Information | Home |
Telephone: 44 (1455) 285520 |
Please E-Mail
your comments and suggestions to OMEGA
or call us @ 0800 488 488
© Copyright 1999 OMEGA ENGINEERING, INC. ALL RIGHTS RESERVED.